Skip to content

Calibrating the MJK Phix pH probe

Download as PDF

pH sensors must be calibrated at regular intervals to provide an accurate reading. Follow the steps below to calibrate an MJK Phix pH probe while it is connected to a Consibio Logger.

The calibration is performed on the probe itself, using the calibration ring and two buffer solutions. The Logger only needs to supply constant power while you work.

  • Deionized water to rinse the probe.
  • A calibration buffer solution with pH 7.0.
  • A calibration buffer solution with pH 4.0.
  • A soft cloth that will not scratch the electrode.

During normal operation, the Logger powers the sensor up and down with every measurement to preserve battery. That interferes with the calibration procedure, so put the Logger into a configuration where it stays online:

  1. Open the device in Consibio Cloud, set the Transmission interval to 0 seconds and select Save. At 0s, the device attempts to stay online indefinitely.

    The device settings in Consibio Cloud with the transmission interval fields set to 0 days, 0 hours, 0 minutes and 0 seconds, and the Save button highlighted

  2. Press Restart shortly on the Logger to force a restart and an immediate connection, so the new setting is synchronized to the device right away.

    The inside of the Logger with a callout pointing at the Restart button, reading “Press Restart shortly”

  3. Wait until Configuration is listed as In sync. in Consibio Cloud. This can take a couple of minutes.

    The device overview in Consibio Cloud with a callout pointing at the Configuration row, reading “Wait until the configuration is listed as In sync.”

  4. Wait at least 10 minutes for the sensor to stabilize.

For more about the interval settings, see Logger WD-68.

  1. Clean the sensor carefully in deionized water, then dry the electrode with a soft cloth.

  2. Turn the calibration ring to the 7 indicator position and submerge the probe in the pH 7 buffer solution.

    The MJK Phix probe submerged in a beaker of pH 7 buffer, with the calibration ring turned so the red marker points at the 7 position

  3. The sensor begins to blink 3 times every 5 seconds.

  4. Wait until the sensor blinks 2 times every 5 seconds. The first calibration value is now saved.

  1. Take the sensor out of the buffer solution and dry the electrode with a soft cloth.

  2. Turn the calibration ring from the 7 to the 4 indicator position and submerge the probe in the pH 4 buffer solution.

    The MJK Phix probe submerged in a beaker of pH 4 buffer, with the calibration ring turned so the red marker points at the 4 position

  3. The sensor begins to blink 3 times every 5 seconds.

  4. Wait until the sensor blinks 2 times every 5 seconds. The second calibration value is now saved.

Turn the calibration ring to the M (Measure) indicator position.

The MJK Phix probe out of solution, with the calibration ring turned so the red marker points at the M position

The sensor is now calibrated.

Set the Transmission interval back to the value you want the Logger to run with, and select Save.

The device settings in Consibio Cloud with the transmission interval fields and the Save button highlighted

If you have any questions, reach out to us at support@consibio.com.